 'Spoiler Gradient Pulse' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Spoiler Gradient Pulse' found in 1 term [] and 7 definitions [], (+ 6 Boolean[] results
previous     11 - 14 (of 14)     
Result Pages : [1]  [2]  [3]
MRI Resources 
Breast Implant - Contrast Enhanced MRI - Mobile MRI - Homepages - Guidance - MRI Physics
 
Spoiled Steady State Acquisition Rewinded Gradient EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(SARGE) Spoiled GRE sequences use a spoiler gradient on the slice select axis to destroy any remaining transverse magnetization after the readout gradient, with the result of short repetition times. This type of sequences use semi-random changes in the phase of RF pulses to produce a spatially independent phase shift.

See also Gradient Echo Sequence.
spacer
MRI Resources 
Safety Training - Pregnancy - Stent - Process Analysis - Spectroscopy - PACS
 
FID Signal ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
FID (with spin echo) signal
DESCRIPTION
Line across the center of the image
REASON
Combination of problems
HELP
Call the service
A combination of B1 inhomogeneity, poor slice profile, and insufficient spoiler gradients between the refocusing pulse and the readout interval of a spin echo sequence results in a FID signal being detected along with the echo.
Since the FID is not phase encoded (normally the phase encoding occurs before the refocusing pulse), it is not dispersed along the phase encoding axis, but appears as a line across the center of the image.
mri safety guidance
Image Guidance
If the problem persists, it must be addressed by a service representative.
spacer

• View the DATABASE results for 'FID Signal Artifact' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
MRI Resources 
Hospitals - MRI Technician and Technologist Career - Jobs pool - Service and Support - Contrast Enhanced MRI - MRI Technician and Technologist Schools
 
Saturation Pulses
 
Sequence of RF (and gradient) pulses designed to produce saturation, typically in a selected region or set of regions, most often by the use of selective excitation followed by a spoiler pulse. Similar to some spectral suppression techniques. Can be used to reduce signal from flowing blood by saturating regions upstream from region being imaged.
spacer

• View the DATABASE results for 'Saturation Pulses' (7).Open this link in a new window

MRI Resources 
Process Analysis - Mobile MRI - Universities - Collections - Open Directory Project - Service and Support
 
Spoiled Gradient Echo SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Spoiled gradient echo sequences use a spoiler gradient on the slice select axis during the end module to destroy any remaining transverse magnetization after the readout gradient, which is the case for short repetition times.
As a result, only z-magnetization remains during a subsequent excitation. This types of sequences use semi-random changes in the phase of radio frequency pulses to produce a spatially independent phase shift.
Companies use different acronyms to describe certain techniques.

Different terms for these gradient echo pulse sequences:
CE-FFE-T1 Contrast Enhanced Fast Field Echo with T1 Weighting,
GFE Gradient Field Echo,
FLASH Fast Low Angle Shot,
PS Partial Saturation,
RF spoiled FAST RF Spoiled Fourier Acquired Steady State Technique,
RSSARGE Radio Frequency Spoiled Steady State Acquisition Rewound Gradient Echo
S-GRE Spoiled Gradient Echo,
SHORT Short Repetition Techniques,
SPGR Spoiled Gradient Recalled (spoiled GRASS),
STAGE T1W T1 weighted Small Tip Angle Gradient Echo,
T1-FAST T1 weighted Fourier Acquired Steady State Technique,
T1-FFE T1 weighted Fast Field Echo.
In this context, 'contrast enhanced' refers to the pulse sequence, it does not mean enhancement with a contrast agent.
spacer

• View the DATABASE results for 'Spoiled Gradient Echo Sequence' (11).Open this link in a new window

 
Further Reading:
  News & More:
3-D VOLUMETRIC IMAGING FOR STEREOTACTIC LESIONAL AND DEEP BRAIN STIMULATION SURGERY
Cutting Edge Imaging of THE Spine
February 2007   by www.pubmedcentral.nih.gov    
MRI Resources 
Distributors - General - Journals - Pediatric and Fetal MRI - MRI Technician and Technologist Jobs - Pacemaker
 
previous      11 - 14 (of 14)     
Result Pages : [1]  [2]  [3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 6 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]